ICP-AES方法线性范围宽,相关系数在0.9995以上,相对标准偏差在1%以下,最低检出限为0.0003%~0.0201%,回收率为90~123%。采用ICP-AES方法测定凹土中金属元素时, 操作简单、快速

摘 要:利用电感耦合等离子体原子发射光谱法(ICP-AES)测定凹土中多种金属离子的含量。将凹凸棒黏土经研磨筛分并加稀盐酸加热消解后配制成溶液供ICP-AES分析。实验结果表明:ICP-AES方法线性范围宽,相关系数在0.9995以上,相对标准偏差在1%以下,最低检出限为0.0003%~0.0201%,回收率为90~123%。采用ICP-AES方法测定凹土中金属元素时, 操作简单、快速,化学干扰少,测定结果准确,精密度好,可以应用于ATP中多种金属元素的含量测定。55298

毕业论文关键词:凹凸棒黏土,电感耦合等离子体发射光谱,金属元素,含量测定

Abstract: Multiple metal elements in attapulgite clay were determined by ICP-AES. By grinding, sieving and hydrochloric acid heating digestion, the attapulgite clay were matched into solution for ICP-AES analysis .Experimental results showed that ICP-AES method had wide linear range, correlation coefficients were above 0.9995, the relative standard deviation was below 1%,the minimum detection limit ranged from 0.0003% to 0.0201% ,the recovery was from 90% to 123%. In determination of multiple metal elements in attapulgite clay ,ICP-AES method is simple, rapid, less chemical interference, measurement results are accurate, precision, and can be applied in the determination of ATP various metal elements.

Keyword: attapulgite clay, ICP-AES, metal elements, determination

目录

1 前言 4

2 实验部分 5

2.1主要仪器、试剂、样品 5

2.2仪器工作参数 5

2.3 实验方法 6

3结果与讨论 6

3.1样品消解 6

3.2消解条件选择 6

3.2.1消解酸的确定 6

3.2.2 HCl用量的确定 7

3.2.3消解时间的确定 8

3.3 仪器工作参数的选择 9

3.3.1功率的选择 9

3.3.2 雾化器压力的选择 11

3.3.3 元素分析线的选择 12

3.4光谱干扰及其校正 14

3.5 标准溶液配制及其标准曲线 14

3.5.1 标液配制 14

3.5.2 标准曲线 14

3.6 方法检出限 17

3.7  方法精密度 18

3.8 回收率及基体干扰实验 19

3.8.1标液配制 19

3.8.2回收率实验 19

3.8.3基体干扰实验 20

3.9样品的测定 22

结 论 24

参考文献 25

致谢 26

1 前言

凹凸棒粘土(以下简称凹土,ATP)是指以凹凸棒粘土为主要矿物成份的一种稀少的天然非金属粘土矿物,在矿物学上隶属于海泡石族。凹凸棒粘土的分子式为Mg2.074Al1.026(Si4O10.48)2(H2O)10.6,表1为江苏盱眙天然凹凸棒粘土的化学成分[1]。

表1 江苏盱眙天然凹凸棒粘土的化学成分

Compotent SiO2 凹土中Mg、Al、Fe等十二种元素测定方法研究:http://www.youerw.com/huaxue/lunwen_59587.html